Page 29
Mounting and Installation
scanCONTROL 29xx
5.2.2
Power Supply
Connector
.
Range: 11 V – 30 V (rated value 24 V) DC; load maximal 500 mA
The operating voltage is protected against reverse polarity.
The cable shield is connected with the connector housing and should be connected to the protective
conductor PE of the power supply.
The shielded multifunction cable PC2600/2900-x is recommended.
The operating voltage for the scanCONTROL 29xx measuring device should come from a 24 V power
supply, which is only used for measuring equipment and not simultaneously for drives, contactors or
similar pulse interference sources. Use a power supply with galvanic isolation.
Connect the multifunc-
tion socket only when
switched off the power
supply is switched off.
> Damage to or
destruction of the
sensor