Mounting and Installation
scanCONTROL 29xx
The multifunction socket can be used with either of the following configurations:
Configuration
In1
In2
In3
0
Encoder with index, positive edge works with the index
1
N
A
B
1
Encoder without index, additionally external trigger possible
1
Trigger
A
B
2
External trigger
Trigger
3
External trigger, load up to 4 user modes
Trigger
Mode Bit 0 Mode Bit 1
4
Load up to 8 user modes
Mode Bit 0 Mode Bit 1 Mode Bit 2
5
Transmit in time stamp, (only 2910/2960)
Bit 0
Bit 1
Bit 2
Signal level (switching level):
The signal levels are switchable for all switching inputs together via software between LLL (low-voltage-, TTL
logic) and HLL (high-voltage-, HTL logic):
- LLL level: Low 0 V… 0.8 V, High 2.4 V… 5 V, internal pull-up 10 kOhm to 5 V
- HLL level: Low 0 V… 3 V, High 11 V… 24 V (permitted up to 30 V), internal pull-up 10 kOhm to 24 V
- Pulse duration:
≥
5 μs
i
Use a shielded cable with twisted wires, especially the recommended connection cable PC2600/2900-x
from the accessories, see Chap.
Connect the cable shield with the potential equalization PE or the connector housing.
1) The encoder input counts each edge. Encoder output typically 4 edges per encoder step.
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......