Page 58
Error Codes
scanCONTROL 29xx
11.
Error Codes
(— LED
state
lights for a long time lang, • LED
state
lights briefly)
Flashing
sequence
Cause
Remedy
Notes
Group: Loading / saving configuration
..
2x short
Mode not found.
Select different one.
Only previously stored
modes can be called up.
..-
2x short, 1x long
White error flash
Contact manufacturer, return
sensor.
Should not occur in normal
operation.
…
3x short
Flash full
None, contact manufacturer.
Should not occur in normal
operation.
….
4x short
Loading suppressed due to
active data transmission.
Stop active data transmission.
Prevents PC software
crashes.
Group: Data processing and transmission
--
2x long
Data overflow in the sensor
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired; ex-
posure time can be longer
than expected.
--.
2x long, 1x short
Data overflow during receipt
of the data from the sensor.
Select smaller measuring field,
reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired.
--..
2x long, 2x short
Data overflow for serial port
RS422
Reduce profile frequency, se-
lect less complex measuring
program.
Data can be impaired.
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......