Page 28
Mounting and Installation
scanCONTROL 29xx
5.2
Connections
5.2.1
General
1 2
1 Ethernet socket
2 Multifunction socket (Power supply, IO)
Fig. 17 Output sockets arrangement
Designation Sensor connector
Pin
Cable color
PC2600/2900-x
Notes
Connection view
+Ub
9
red
+ 11 V - 30 V DC (rated
value 24 V); max. 500 mA
2
3
4
5
6
7
8
9
1
10
11
12
screw connector,
solder pin side
GND
2
blue
0 V
+Laser on/off
3
white
optional
-Laser on/off
1
brown
RS422
12
red-blue
RS422
input or output
/RS422
11
gray-pink
In1
6
yellow
Switching input In1
GND-In1
4
green
Ground connection In1
In2
5
pink
Switching input In2
GND-In2
8
gray
Ground connection In2
In3
10
purple
Switching input In3
GND-In3
7
black
Ground connection In3
Screen
Housing
black
DC-insulated from GND
Fig. 18 Assignment of the multifunction socket on scanCONTROL 29xx
GND: Galvanically isolated from IN1, 2, 3, RS422, Laser on/off
Laser on/off: Input galvanically isolated from GND, IN1…3, RS422
IN1, IN2, IN3, RS422: Inputs galvanically isolated from GND and Laser on/off
Summary of Contents for 4350006.71
Page 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Page 95: ...Operating Instructions scanCONTROL 29xx ...
Page 100: ......
Page 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Page 193: ......
Page 194: ......
Page 229: ......