MTS600 Instruction Manual
83
②
Input
“
1234
”
and press [OK], enter <Instr.
Adjustment > menu.
③
Press [F2] to select <Tilt sensor offset>.
④
Accurately collimate a target in Face1, then press
[OK].
l
Press [CE] to cancel the last operation and redo it.
⑤
Loosen the horizontal clamp, rotate the top of the
instrument through 180
°
, accurately collimate the
same target in Face 2. If there
’
s no wrong with
operation, the offset value of tile sensor displayed..
⑥
Press [OK] to accept the new values for correction
and return to <Instr. Adjustment > menu.
。
⑦
Press [ESC] twice ,return to basic measurement mode. Follow the steps
mentioned above, check tilt compensator zero point error again.
⑧
If the error exceeds the desired range yet, please check if there
’
s a wrong with
the operation of adjustment. If not, please send the instrument back to factory for
repair.
Instr.Adjustment
F1-V.ang. index offset
F2-Tile sensor offset
F3-Instrument constant
Tilt ang 0
°
00
′
12
″
H.ang(
R
) 45
°
23
′
07
″
<
一
>Collimate in face 1?
CE
OK
Tilt ang 0
°
00
′
12
″
H.ang(
R
) 45
°
23
′
07
″
<
二
> Collimate in face 2?
CE
OK
Tilt Sensor Offset
Offs. -0
°
00
′
36
″
Accept new value?
CE
OK
Instr.Adjustment
F1-V.ang. index offset
F2-Tile sensor offset
F3-Instrument constant