Instruction for use URYXXON
®
500
24
Instruction for use URYXXON
®
500 V 2.1 / 01.23
Seq.No. range
Selects a specific sequence number range. After you press the
Seq.No., two empty edit‑buttons with four “0” appear. To enter
the sequence numbers, touch the buttons and enter the numbers
via the input‑menu
abc
. Press
, to confirm and save your
changes. If you leave the menu by pressing
, none of your
changes will be adopted.
Search memory
Date
Filter
Action
All
Seq.No.
range
Display
0000
0000
-
Not printed
Selects all results that have not yet been printed.
Not sent
Selects all results that have not yet been sent.
Positive
Selects all results with at least one positive value.
Sediment
Selects all results that already contain sediment information.
Error
Selects all results with error messages.
10.2.3 Action
The activity‑button
Action
determines which action will be
performed for the results that have been specified before using the
date and filter functions. Press the button and all action options
appear. Press
, to confirm and save your changes. If you
leave the menu by pressing
, none of your changes will be
adopted.
Search memory
Date
Filter
Action
Display
Send
Print list
Delete
Display
Displays the selected results.
Prints the selected results.
Send
Sends the selected results to a PC / LIS.
Print list
Prints the selected patient IDs and sequence numbers only.
Delete (Filter selection)
Deletes the results selected with the search function.
Delete (complete memory)
The activity‑button
Delete
within the memory‑menu deletes the
entire memory. To do so, press the button and confirm the following
security question.
NOTICE
Saved data of control measurements will not be deleted with
this function.
NOTICE
Make sure the memory does not contain anymore relevant data,
before you delete results. Once data has been deleted from
the instrument memory, it cannot be recovered. We therefore
recommend sending the data to a PC first.
11. Check mode
The
Check mode
menu allows you to access the quality control
functions
(see chapter 14. Quality control, page 28)
.
Check mode
Settings
Enter sediment
Memory
Check mode
Stored data
Internal check
Control strip
Control
measurement
11.1 Internal check
The self‑test is repeated and the measuring values are displayed.
11.2 Control measurement
This button initiates a control measurement with artificial urine.
11.3 Control strip
This button initiates the measurement of a colored control strip
as an inspection equipment monitoring tool. For ordering of the
colored control strips contact MACHEREY‑NAGEL or your local
distributor.
11.4 Stored data
This button allows you to access the saved results of the last 20
control measurements. Press the button to display the results. You
can either print or send the data.