Instruction for use URYXXON
®
500
10
Instruction for use URYXXON
®
500 V 2.1 / 01.23
3. Instrument set-up
3.1 Package content
Open the box carefully with a sharp instrument. Make sure not to
damage the content. Take out the instrument and all other parts
carefully. Check the package, instrument and all accessories for
visible damage. If a part is damaged, contact your distributor or the
technical support
(see chapter 17.13 Contact, page 34)
The following picture shows the complete package content. Check
the shipment for completeness. If a part is missing, contact your
distributor or the technical support as well.
NOTICE
Keep the original box and the packaging material from the initial
shipment to optimally protect the instrument in case of a return
shipment.
3.2 Set-up location
Only use the instrument in a suitable location. It should be
positioned on a dry, clean, level and horizontal surface.
The operating temperature is between 4 – 30 °C, while the optimal
temperature range for the instrument is between 18 – 25 °C. For
optimal performance, the humidity should be between 20 – 80 %.
The instrument must not be used above 2000 m.
Make sure to protect the instrument from drastic changes in
temperature. Avoid positioning the instrument close to windows,
hot plates, ovens etc. To ensure correct functioning and reliable
results, the instrument should not be exposed to direct sunlight or
other bright, focused light sources.
NOTICE
When operating outside these limits, the correct functioning of
the device is not guaranteed.
a
URYXXON
®
500
b
Transport rack
c
Thermal printer paper rolls
d
Instruction for use
e
Serial cable RS232
f
Power cable
g
Power adapter