PERFORMANCE VERIFICATION TESTS
KI23400/KI27400 UM-2
-
90 -
Model:
Instrument Connector:
APC [ ] PC [ ]
Report No.:
Date:
Calibration Test
Wavelength
Acceptable Range
Measurement Results
1310 nm
-0.5 to +0.5 dB
dB R
1550 nm
-0.5 to +0.5 dB
dB R
dB R
dB R
dB R
Fiber Type
:
Singlemode [ ] Multimode [ ]
Linearity and Range Test
No
Attenuator setting
Meter Reading
Linearity Calculation
Acceptable Linearity Range
1
1
5.00 dB
-
2
15.00 dB
No 2
–
No 1 =
9.50 ~ 10.50
3
20.00 / 25.00 dB
No 3
–
No 2 =
4.50 ~ 5.50 / 9.50 ~ 10.50
4
35.00 dB
No 4
–
No 3 =
9.50 ~ 10.50
Note1: The linearity range here is typically limited by attenuator linearity, rather than instrument performance
Table 7: Test Record for Two Way /One Way Autotest Performance Verification Tests, KI23400/27400 Series