PERFORMANCE VERIFICATION TESTS
KI23400/KI27400 UM-2
-
87 -
Model:
Report No.:
Date:
Total Uncertainty / Accuracy Test
Test Wavelength = nm
Setting
Number
Power Meter
reference value
Attenuator
Setting
Minimum specification
(-0.3 dB of Ref.)
DUT
measurement results
Maximum specification
(+0.3 dB of Ref.)
1.
(~-10.00 dBm)
dB
(~-10.30 dBm)
dBm
(~-9.70 dBm)
2.
(~-20.00 dBm)
dB
(~-20.30 dBm)
dBm
(~-19.70 dBm)
3.
(~-30.00 dBm)
dB
(~-30.30 dBm)
dBm
(~-29.70 dBm)
4.
(~-40.00 dBm)
dB
(~-40.30 dBm)
dBm
(~-39.70 dBm)
5.
(~-50.00 dBm)
dB
(~-50.30 dBm)
dBm
(~-49.70 dBm)
Measurement Uncertainty
dB
Note 1
: This is for the KI2600-InGaAs. For the H series instruments, increase by ±0.2 dB.
Table 4: Accuracy Test Record for KI273400/27400 Series, Power Meter section