7-6 Keysight N5290/1A Service Guide
Repair and Replacement Procedures
Removing the N5292A Option 200/400 from the System
7-
Removing the N5292A Option 200/400 from the System
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— 5/16-in torque wrench (set to 8 in-lbs)
— 5/16-in torque wrench (set to 10 in-lbs)
— 5/16-in torque wrench (set to 21 in-lbs)
— 14 mm open-end wrench (set to 4 in-lbs)
— 20 mm open-end wrench (set to 8 in-lbs)
Figure 7-2
To avoid damaging your RF ports 1/2/3/4, always use two wrenches
IMPORTANT!
The N52xxB interconnect kits contain 2.4 mm to 3.5 mm cables. It is
important to connect the 2.4 mm end to the PNA and to connect the
3.5 mm end to the N5292A test set controller. Improper connections will
result in damage to your instruments.
It is important to use two wrenches when disconnecting or connecting the
semirigid cables to the 1/2/3/4 port 1.85 mm outputs front ports of the
PNA to avoid stressing the connectors and causing damage to your PNA
(i.e., the 2.4 mm cable end is compatible with the 1.85 mm connector of
the PNA).
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...