Keysight N5290/1A Service Guide 3-33
System Tests and Adjustments
N5292A Functional Verification Procedure
When a defective cable or connector has been replaced and the test set
has been repaired, the following tests are required:
—
Test set and N5290A/91 A system calibration
— Insertion Loss/Match-S-parameters (verify PNA Cal has been
—
LO Verify (Run after Insertion Loss/Match are verified.)
IMPORTANT!
LO power verification is only required on the port that the
connector or and cabling was replaced.