Keysight N5290/1A Service Guide 3-15
System Tests and Adjustments
The N5290/1A Millimeter Test System Preliminary Check
If the Head Check Fails
1.
Clean the test ports, shorts and adapters. Torque to specification. Repeat
the check.
2.
If the head check still fails and the failure is one port only: Then swap the
suspect head onto one of the other good ports (i.e., requires running the
N5290/1A system adjustments. Refer to
the problem follows the head, return the head to the factory for repair.
Refer to
“Contacting Keysight” on page 2-7
.
3.
If the failure is on all of the ports, swap in a known good head to verify the
problem is not the test set.
4.
If the check still fails, and the test set is OK, the head should be repaired or
replaced and is a return-to-factory repair. For more on troubleshooting the
millimeter heads refer to
“Procedure for identifying a faulty millimeter
.
5.
If the head(s) are OK, but the PNA is suspected as the problem, remove all
cables from the PNA to the test set and run the PNA Operator’s Check.
https://literature.cdn.keysight.com/litweb/pdf/E8356-
90001.pdf
.
If the PNA fails the Operator’s Check, refer to the PNA’s Service Guide for
https://www.keysight.com/find/pna
or to
.
6.
If the PNA passes the Operator’s Check, and the heads are OK, then the
N5292A is faulty. Refer to
“N5292A Functional Verification Procedure” on
.
If the PNA fails the Operator’s Check, repair the PNA or return the PNA to
the factory for repair. Refer to the Service Guide for your PNA or to
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...