7-8 Keysight N5290/1A Service Guide
Repair and Replacement Procedures
Removing the Cover
7-
Removing the Cover
Tools Required
—
T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
Removing the Outer Cover
Refer to
for this procedure.
1.
Disconnect the power cord.
2.
Remove the strap handles (item
④
) by loosening the screws (item
⑤
),
with a T-20 TORX driver, on both ends until the handle is free of the test
set.
3.
Remove the foot locks (item
⑦
) from the four bottom feet (item
⑥
) and
then remove the four bottom feet from the outer cover.
4.
Remove the four rear panel feet (item
➁
) by removing the center screws
(item
③
) with a T-20 TORX driver.
5.
Slide the outer cover (item
①
) toward the rear of the test set and remove
it.
Replacement Procedure
1.
On the bottom side of the N5292A, carefully fold or push down the ribbon
cables and wires so they can’t be pinched between the hardware and the
outer cover. Ribbon cables and wires must never be positioned on top of
hardware.
2.
Reverse the order of the removal procedures above.
This procedure is best performed with the test set resting on its front
handles in the vertical position. Do not place the test set on its front panel
without the handles. This will damage the front panel assemblies.
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...