Keysight N5290/1A Service Guide 3-45
System Tests and Adjustments
Test Management Environment (TME) Performance Tests (N5290/1 A Millimeter
System Only)
Test Management Environment (TME) Performance Tests (N5290/1
A Millimeter System Only)
The performance tests verify the electrical performance of your N5290/1 A
Millimeter System. For more information, refer to the TME webpage at
http://cal.software.keysight.com
http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm
.
This section lists the test equipment needed to run performance tests. Models
for each test equipment category are given in order of preference. Alternate
models are also listed but should be considered a lower priority. All equipment
is Keysight, Agilent, or Hewlett Packard unless otherwise specified. Refer to
each individual test for specific setup configurations. The computer and all test
equipment are connected to the unit under test (UUT) using GPIB.
Guidelines for Ordering Power-Sensor Calibration Service
Use the following guidelines to ensure power sensors used by the N7800A
application receive the correct calibration service:
1.
Order Option 1A7 when ordering new power sensors that will be used as
working standards in N7800A TME calibrations.
2.
Order the Keysight Cal+ Measurement Uncer GB for power sensor
recalibration.
3.
When ordering periodic calibration for instruments used as lab standards
in the N7800A software, we recommend using "Keysight calib
uncertainties" for power sensors, and "Keysight calib uncertainties
+ guardbanding" for all other items (please visit Selecting the Right
Calibration Services). The N7800A software incorporates the ISO GUM
Uncertainty in point-to-point uncertainty calculations. The special
"H-series" calibration options in this table provide lower measurement
uncertainties through use of direct comparison to devices directly
Requires a PC controller to run the (TME) Performance Tests.
For a list of the equipment required for your system, refer to the N7840A
Calibration Application Help for PNA-X Series Network Analyzer Help and
to the "Test Equipment" topic.
http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm
Power sensor calibrations services:
"Keysight Cal + Measurement Uncer GB" that follows ISO
17025/ILAC-G8 standards.
(
http://cal.software.keysight.com/PNA/Help/N7840AWebHelp.htm
)
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...