3-14 Keysight N5290/1A Service Guide
System Tests and Adjustments
The N5290/1A Millimeter Test System Preliminary Check
3-
Figure 3-8
Lifting the cable at the bend and watching for trace spikes/instability
7.
Gently extend the head (exercising care to not overextend) and rotate side
to side (less than 90 degree) making sure traces are relatively stable and
still tracking each other. Refer to
.
Figure 3-9
Gently extending the head and exercise carefully so as not to overextend
8.
Repeat steps
through
as required for suspected faulty ports.