3-32 Keysight N5290/1A Service Guide
System Tests and Adjustments
N5292A Functional Verification Procedure
3-
Figure 3-20
ECal Connected to Ports 1 and 2 for 2-port Calibration and Insertion
Loss/Transmission baseline integrity Measurements
3.
Verify Test Set S-Parameters (Using the Launch Test application): Fixtures
that are used are shown in
along with a typical
ET setup. This will check for potential failures between the test set's
adapters and port connectors all the way to the TSMB in the following
cabling paths:
—
Low band path Transmission and Match
—
REF IF path transmission and match
—
TEST IF path transmission and match
—
RF path transmission and match
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...