Keysight N5290/1A Service Guide 7-29
Repair and Replacement Procedures
Removing and Replacing the Subpanel PCA Sensor Boards
Removing and Replacing the Subpanel PCA Sensor Boards
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— 5/16 inch open-end torque wrench (set to 10 in-lb)
— ESD grounding wrist strap
Removal Procedure
1.
Disconnect the power cord.
2.
Remove the outer covers. Refer to
“Removing the Cover” on page 7-8
.
3.
Remove the front panel. Refer to
“Front Panel Assembly, Front Side, All
and
“Front Panel Assembly, Back Side, All Options”
4.
Disconnect the power switch cable from the power switch board. Refer to
and to
.
5.
Remove the front panel and set aside test set. For the location of the
screws, see
.
6.
Subpanel: Remove the cables for the port(s) that require the port sensor
PCA to be replaced by referring to
“Removing and Replacing the Front
Panel and Subpanel Assemblies” on page 7-10
. See also
Assembly, Front Side, All Options” on page 6-9
Assembly, Back Side, All Options” on page 6-13
.
7.
Remove the screw securing the port sensor PCA to the subpanel. Refer to
.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Perform the post-repair adjustments, verifications, and performance tests
that pertain to this removal procedure. Refer to
.
Be careful not to damage the center pins of the semirigid cables. Some
flexing of the cables is necessary to remove the assembly. However, do not
over-bend the semirigid cables.
Summary of Contents for N5222B
Page 3: ...3 ...
Page 4: ...4 ...
Page 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Page 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Page 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Page 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Page 225: ...Installation Note Xxxxx xxxxx 3 ...