Keysight N5290/1A Service Guide 3-13
System Tests and Adjustments
The N5290/1A Millimeter Test System Preliminary Check
g.
If your instruments are not warmed up for two hours, repeat step
Else, skip to next step.
5.
Both traces should track each other and look relatively flat. Refer to
.
Figure 3-7
Typical Cable Return Loss Response (port 2 trace is displayed below)
6.
While observing the traces, gently lift the cable (at or near where the
cable bends) up and down a few times. Traces should remain stable and
free from any major spikes or holes (
≤
0.1 dB p–p). Refer to
.