5
4 Disable all analyzers except the one being tested. This simplifies the instructions and makes
module initialization faster.
a
Select the
Overview
tab at the bottom of the main window.
5 Set up the bus:
a
In the
Overview
window, select
Setup -> Bus/Signal...
from the module’s drop-down menu.
b
Scroll if necessary to view the pods you are testing.
c
Verify that the pod threshold buttons say “
Threshold: Differential
”. If they don’t, make sure the
correct probes (E5379A) are attached to pods 1 and 2. The threshold is set to Differential
automatically when E5379A probes are attached.