Keysight 16860 Series Portable Logic Analyzer Service Guide
69
Testing 16860 Performance
3
4 Open the Sampling tab in the Analyzer Setup dialog by clicking the
Sampling Setup
Icon.
5 In the clock assignment area, set Pod 1 Clock to
Falling Edge
.
6 Adjust the sample position using the procedure described earlier by clicking the
Eye Scan: Sample
Position and Thresholds
button.
7 Close the dialog windows by clicking
OK
.
Single Clock Falling Edge - Pod 1 Data - Measuring Maximum Data Rate
1 Select the
Run Repetitive
icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
3 If the clock rate is too high, this error message may occur.