3
Testing 16860 Performance
50
Keysight 16860 Series Portable Logic Analyzer Service Guide
13 After setting to suggested, all bits should be centered in their own eye near -.5 ns.
14 Close Eye Scan by clicking
OK
.
15 Close the Sampling dialog by clicking
OK
.
Pod 1 Clock - Measuring Maximum Data Rate
After setting the sample positions, the data capture can now be verified at the frequency found in the
clock rate section. The configuration file that was loaded earlier with markers is set up to verify the
data pattern. If bad data is found, the marker will cause the repetitive run to stop.
1 Select the
Run Repetitive
icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...