Keysight 16860 Series Portable Logic Analyzer Service Guide
43
Testing 16860 Performance
3
8 In the
Save As
dialog, type the file name
SingleClock_NoMarker.ala
. For file options, select
Setup
Only
. Then click
Save
.
Create Configuration File 4 - Single Clock with Marker
1 Start with the
MultipleClocks_WithMarker
configuration file. Select
File
and then
Open
. In the Open
dialog select the file then open. If asked about saving the current configuration, click
No
.
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods. The Pod 1 leads should be connected to the generator as described
above.
3 Open the
Analyzer Setup Sampling
tab.
4 In the
State Options
section, change the Sampling options to
Single Clock
.
5 In the clock selection area, change the Pod 1 clock to
Both Edges
.
6 Close the dialog by clicking
OK
.
7 Save the configuration by selecting
File
then
Save As …
.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...