5
Troubleshooting
100
Keysight 16860 Series Portable Logic Analyzers Service Guide
d
Channels 7 through 0 are already assigned by default. Assign pod 2 channels 15 through 0 and
pod 1 channels 15 through 8 by clicking and dragging from the left-most channel box to the
right-most channel box. Your display should look like the lower picture when you are done.
6 Select the State sampling mode and set the State Clock options:
a
Select the
Sampling
tab of the Analyzer Setup window.
b
Select
State - Synchronous Sampling
.
c
For State Clock, select
CLK1 Clock
and
Both edges
.
7 Set the trigger position and acquisition memory depth:
a
Set the Trigger Position to
100% Poststore
.
b
Set the
Acquisition Depth
to
8K
.