Keysight 16860 Series Portable Logic Analyzers Service Guide
87
Troubleshooting
5
To use the logic acquisition troubleshooting flowcharts
Flowcharts are the primary tool used to isolate defective assemblies. The flowcharts refer to other
tests to help isolate the trouble. The circled numbers on the charts indicate connections with the
other flowchart. Start your troubleshooting at the top of the first flowchart.
If the instrument still doesn't work correctly after completing all the procedures described in the
flowchart, return it to Keysight Technologies for repair. Be sure to include a note describing the
problem in detail.