Keysight 16860 Series Portable Logic Analyzer Service Guide
25
Testing 16860 Performance
3
Perform the Self-Tests
Power the 16860 series analyzer on and start the Logic Analyzer application.
1 Before performing the self- tests, disconnect all probes from the logic analyzer.
2 Select
Help
->
Self-Test..
. from the main menu. The
Analysis System Self Tests
window appears.
3 From the
Select suites
list, select the option displayed for 16860. Then, select
All
from the
Select
tests
list.
4 Select
Start
to start a complete module self-test. The progress of the self tests is displayed in the
Results
area of the window.
5 When the self-tests are complete, check the
Results
window to ensure that the Result Summary
says that all tests passed. If all tests did not pass, refer to
6 Select the
Close
button to close the
Analysis System Self Tests
window.
7 If all self- tests pass, then record “PASS” in the “Logic Analysis System Self-Tests” section of the
Performance Test Record (
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...