Keysight 16860 Series Portable Logic Analyzers Service Guide
103
Troubleshooting
5
12 Repeat the cable test for pods 3 and 4:
a
Connect the logic analyzer’s pod 3 cable to the stimulus board’s pod 4 connector.
b
Connect the logic analyzer’s pod 4 cable to the stimulus board’s pod 5 connector.
c
In the
Buses/Signals
tab of the Analyzer Setup window, assign the pod 3 and 4 channels to “My
Bus 1”.
d
In the
Sampling
tab of the Analyzer Setup window, for State Clock, select
Both edges
.
e
Adjust sampling positions.
f
Click the
Run
icon
. In the
Listing
window, check at least 256 samples for failures; if
necessary, verify any failures by swapping the E5379A probes.
13 Repeat the cable test (step12) for pod pairs 5 and 6 and 7 and 8 connecting the odd pod cable to
the stimulus board’s pod 4 connector and the even pod cable to the stimulus board’s pod 5
connector:
Return to the troubleshooting flowchart.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...