Testing 16860 Performance
3
Pod 4 Clock - Maximum Clock Rate
1 Load the
MultipleClocks_NoMarker.ala
configuration file. If asked to save the current configuration
file, click
No
.
2 Disconnect the U4203A Flying Lead Probe Set from channels 1 & 2 of the 81134A pulse
generator output (Bits 2, 6, 10, 14) and clock leads.
3 Connect the probe set from Pod 4 of logic analyzer to the pulse generator channels 1 & 2 output.
• Clock to Channel 1 Output
• Clock (NOT) Channel 1 Output (not)
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
4 Open the
Sampling
tab in the Analyzer Setup dialog by clicking the
Sampling Setup
Icon.
5 In the clock assignment area, set Pod 1 Clock to
Don't Care
and set Pod 4 Clock to
Both Edges
.
6 Close the dialog window by clicking
OK
.
7 Set the generator frequency to 357 MHz as a starting point.
8 Click the
Run Repetitive
toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
9 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
11 Without changing the frequency, clear the message by clicking OK. Then click the
Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error, increase the frequency by 1 MHz and run again.
13 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
14 Set the generator to the highest frequency that did not cause an error.
15 Click
Run Repetitive
.
NOTE
This section is only applicable to models: 16862A, 16863A & 16864A.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...