3
Testing 16860 Performance
74
Keysight 16860 Series Portable Logic Analyzer Service Guide
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
4 Open the
Bus/Signal Setup
dialog by clicking the
Setup
Icon in the toolbar.
5 Unassign the data bits from Pod1.
6 Assign bits 2, 6, 10, and 14 of Pod 2.
7 Open the
Sampling
tab.
8 Adjust the sample position using the procedure described earlier by clicking the
Eye Scan: Sample
Position and Thresholds
button. Select the eye closest to -1 ns.
9 Close the dialog windows by clicking
OK
.
Single Clock Both Edges - Pod 2 Data - Measuring Maximum Data Rate
1 Select the
Run Repetitive
icon
. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window appears. The data in the listing window should be
A's & 5's.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...