Keysight 16860 Series Portable Logic Analyzer Service Guide
35
Testing 16860 Performance
3
Test the 16860 series Logic Analyzer
The following sections explain how to measure the maximum state data rate.
1 Record the logic analyzer's model and serial number in the Performance Test Record (see
). Record your work order number (if applicable) and today's date.
2 Record the test equipment information in the “Test Equipment Used” section of the Performance
Test Record.
3 Turn on the frame.
a Plug in the power cord to the power connector on the rear panel of the logic analyzer.
b Connect a key board and a mouse to the frame.
c Press the Standby (Power) button on the front panel of the logic analyzer to power on the logic
analyzer.
While the logic analyzer is booting, observe for the following:
• Ensure all of the installed memory is recognized.
• Any error messages.
• Interrupt of the boot process with or without error message.
4 During initialization, check for any failures.
If an error or an interrupt occurs, refer to the chapter
5 Start the Keysight Logic Analyzer application if it is not started already.
NOTE
Before testing the performance of the logic analyzer, warm-up the logic analyzer and the test
equipment for 30 minutes.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...