Keysight 16860 Series Portable Logic Analyzer Service Guide
57
Testing 16860 Performance
3
4 Unassign the data bits from Pod1.
5 Assign bits 2, 6, 10, and 14 of Pod 3.
6 Open the
Sampling
tab.
7 In the clock assignment area, set Pod 1 Clock to
Don't Care
and set Pod 3 Clock to
Both Edges
.
8 Adjust the sample position using the procedure described earlier by clicking the Eye Scan:
Sample Position and Thresholds button.
9 Close the dialog windows by clicking
OK
.
Pod 3 Clock - Measuring Maximum Data Rate