Keysight 16860 Series Portable Logic Analyzer Service Guide
65
Testing 16860 Performance
3
Single Clock Rising Edge - Pod 1 Data - Setup for Maximum Data Rate
To measure the maximum data rate, the starting frequency will be the frequency found in the section
above. A new configuration file will be used that has a marker setup to verify that the data patterns
are correct. To ensure that the data is sampled correctly, the sample position needs to be set. Eye
Scan is used to set the sample points of all the data to the same eye.
1 On the 81134A, select the channel 2 setup screen. This is the data channel.
2 Set the Channel 2 Freq. Divider to 2.
3 Load the
SingleClock_WithMarker.ala
configuration file. If asked to save the current configuration
file, click
No
.
4 Verify that the Generator is set to the frequency found in the last section as a starting point.
5 Open the Sampling tab in the Analyzer Setup dialog by clicking the
Sampling Setup
Icon.
6 In the clock assignment area, set Pod 1 Clock to
Rising Edge
.
7 Adjust the sample position using the procedure described earlier by clicking the
Eye Scan: Sample
Position and Thresholds
button.
8 Close the dialog windows by clicking
OK
.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...