3
Testing 16860 Performance
48
Keysight 16860 Series Portable Logic Analyzer Service Guide
6 In the dialog, set
Do full time/voltage scan
and
Show time/voltage diagram
.
7 Run Eye Scan by clicking the
Run This Measurement
button.
8 If the scan does not fill the scan area, right click in the scan area and select
Autoscale all
.
The waveform should now fill the Eye Scan area.
9 Move the sample position of all 4 data bits into the Eye between 0 and -1ns. Using the mouse,
grab the vertical lines and move them into the correct eye.