3
Testing 16860 Performance
68
Keysight 16860 Series Portable Logic Analyzer Service Guide
9 Without changing the frequency, clear the message by clicking OK. Then click the
Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
10 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error, increase the frequency by 1 MHz and run again.
11 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
12 Set the generator to the highest frequency that did not cause an error.
13 Click
Run Repetitive
.
14 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
15 Click the
Stop
toolbar button to stop the data acquisition.
16 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Single Clock Falling Edge - Pod 1 Data - Setup for Maximum Data Rate
1 On the 81134A, verify that Channel 2 Freq Divide is set to 2.
2 Load the
SingleClock_WithMarker.ala
configuration file. If asked to save the current configuration
file, click
No
.
3 Verify that the Generator is set to the frequency found in the last section as a starting point.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...