Keysight 16860 Series Portable Logic Analyzer Service Guide
31
Testing 16860 Performance
3
•
Option 700: Test at 700 Mb/s plus 2% (714 Mb/s). Set the generator to half this or 357
MHz.
This includes the frequency uncertainty of the pulse generator, cabling, and a test margin. If
you are using an 81134A pulse generator, the frequency accuracy is ±0.005% of setting.
b Set the rest of the pulse generator parameters to the values shown in the following tables.
Table 4 81134A Pulse Generator Setup
Main
Channel 1
Channel 2
Mode: Pulse/Pattern
Mode: Square ÷ 1
Mode: Square ÷ 1
Freq: set in previous step
Timing
Timing
Clock Internal
Delay Ctrl Input Off
Delay Ctrl Input Off
Delay 0 ps
Delay 0 ps
Pulse Perf: Normal
Pulse Perf: Normal
Deskew: 0 ps
Deskew: 0 ps
Levels: Normal, Custom
Levels: Normal, Custom
Ampl: 450 mV
Ampl: 450 mV
Offset: 0 mV
Offset: 0 mV
Term Voltage: 0 mV
Term Voltage: 0 mV
Limit to current Levels: unselected
Limit to current Levels: unselected
Output: Enable (LED on)
Output: Enable (LED on)
Output: Enable (LED on)
Output: : Enable (LED on)
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...