Keysight 16860 Series Portable Logic Analyzers Service Guide
101
Troubleshooting
5
8 Adjust sampling positions:
a
Click the
Eye Scan: Thresholds and Sample Positions
button. The
Eyescan - Sample Positions and
Threshold Settings
dialog will appear.
b
In the “Buses/Signals” section of the Eyescan - Sample Positions and Threshold Settings
dialog, make sure the check box next to “My Bus 1” is checked.
c
Click
Edit,
select the
Do time scan only
option as the scan dimension and then select the
Lock
Thresholds
checkbox.
d
Select the
Run This Measurement
button in the
Eyescan - Sample Positions and Threshold Settings
dialog.
y
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...