3
Testing 16860 Performance
62
Keysight 16860 Series Portable Logic Analyzer Service Guide
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
be rerun and the test should be rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, select the
Stop
button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For this Pod's Clock, record the generator frequency and the Data Rate in the "Maximum State
Data Rate" section of the
on page 77. Remember that the data rate is
twice the generator frequency.