6
Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies
114
Keysight 16860 Series Portable Logic Analyzer Service Guide
5 Reverse this procedure to install the front panel assembly.
Disassembling the Front Panel
After you have removed the front panel assembly using the procedure described above, you can
disassemble the front panel using the procedure given below.
1 Remove the knob on the front panel by gently pulling the knob forward.
2 Carefully disconnect the backlight cable from the two locations on the front panel highlighted in
the following photo.
CAUTION
Do not remove the backlight cable from the open-faced adhesive panel
unless you need to replace this cable.