6
Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies
126
Keysight 16860 Series Portable Logic Analyzer Service Guide
Removing and Replacing the Power Supply Shroud and Power Supply
Prerequisites
Ensure that you have:
• completed the tasks listed in the
topic.
• removed the cover using the procedure listed in
“Removing and Replacing the Hard Drive,
• removed the front panel using the procedure listed in
“Removing and Replacing the Front
To remove and replace the power supply and its shroud
1 Unclip and then disconnect the Fan power cable as shown in the photo below.
2 Using a Torx T10 screwdriver, remove the four screws that secure the power supply shroud to the
frame. Two of these screws are located inside the frame and remaining two are located outside
the frame. These screws are displayed in the photos shown below.
s
WARNING
Hazardous voltages exist on the power supply. To avoid electrical shock,
disconnect the power from the instrument before performing the following
procedures. After disconnecting the power, wait at least six minutes for the
capacitors on the power supply board to discharge before servicing the
instrument.
Two screws inside the frame
Two screws outside the frame
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...