4-10
Model 2790 SourceMeter
®
Switch System Reference Manual
Bandwidth
Bandwidth specifies the lowest frequency of interest for AC measurements. The RATE
setting determines the bandwidth setting:
•
SLOW — 3Hz to 300kHz
•
MEDium — 30Hz to 300kHz
•
FAST — 300Hz to 300kHz
When the Slow bandwidth (3Hz to 300kHz) is chosen, the signal goes through an analog
RMS converter. The output of the RMS converter goes to a fast (1kHz) sampling A/D and
the RMS value is calculated from 1200 digitized samples (1.2s).
When the Medium bandwidth (30Hz to 300kHz) is chosen, the same circuit is used. How-
ever, only 120 samples (120ms) are needed for an accurate calculation because the analog
RMS converter has turned most of the signal to DC.
In the Fast bandwidth (300Hz to 300kHz), the output of the analog RMS converter (nearly
pure DC at these frequencies) is simply measured at 1 PLC (16.6ms). For remote program-
ming, the integration rate can be set from 0.01 PLC to 50 or 60 PLC.
lists the front panel bandwidth settings for the AC measurement functions. For
remote programming, the FAST, MED, and SLOW annunciators are only lit when condi-
tions in the table are met. In other cases, the annunciators are turned off.
Scanning (7702 only)
When a simple scan using the 7702 module is configured, the present rate or bandwidth
setting will apply to all channels in the scan. When an advanced scan is configured, each
channel can have its own rate or bandwidth setting. Details to configure and run a scan are
provided in
.
For remote programming, the <clist> parameter is used to configure channels for a scan.
Remote programming — rate and bandwidth
Rate and bandwidth commands
The commands to set the integration rate and bandwidth are listed in
. Additional
information on these commands follows the table.
klqb
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...