11-18
Model 2790 SourceMeter
®
Switch System Reference Manual
Figure 11-7
Questionable event register
Condition registers
shows, each status register set (except the Standard Event Register set) has
a condition register. A condition register is a real-time, read-only register that constantly
updates to reflect the present operating conditions of the instrument. For example, while
the Model 2790 is in the idle state, bit B10 (Idle) of the Operation Condition Register will
be set. When the instrument is taken out of idle, bit B10 clears.
The commands to read the condition registers are listed in
. For details on read-
ing registers, see
“Reading registers,” page 11-6
.
Table 11-3
Condition register commands
Command
Description
STATus:OPERation:CONDition?
Read Operation Condition Register.
STATus:MEASurement:CONDition? Read Measurement Condition Register.
STATus:QUEStionable:CONDition? Read Questionable Condition Register.
(B14)
(B15)
OR
Questionable
Condition
Register
Questionable
Event Enable
Register
Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
& = Logical AND
OR = Logical OR
0
(B14)
(B8)
(B15)
0
Warn
Cal
Questionable
Event
Register
(B14)
(B15)
To
Questionable
Summary Bit of
Status (QSB)
Byte Register
(B3 - B0)
Warn
0
Warn
&
&
&
Int2 Int1
(B9)
(B13 - B11) (B10)
Temp
(B4)
(B7 - B5)
(B8)
Cal
(B3 - B0)
Int2 Int1
(B9)
(B13 - B11) (B10)
Temp
(B4)
(B7 - B5)
(B8)
Cal
(B3 - B0)
Int2 Int1
(B9)
(B13 - B11) (B10)
Temp
(B4)
(B7 - B5)
&
&
&
Int2 = Slot 2 Interlock Open
Int1 = Slot 1 Interlock Open
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...