10-12
Model 2790 SourceMeter
®
Switch System Reference Manual
<n>
Numeric value — A numeric value parameter can consist of an NRf num-
ber or one of the following name parameters; DEFault, MINimum,
MAXimum. When the DEFault parameter is used, the instrument is pro-
grammed to the *RST default value. When the MINimum parameter is
used, the instrument is programmed to the lowest allowable value. When
the MAXimum parameter is used, the instrument is programmed to the
largest allowable value. Examples:
ARM:TIMer 0.1 Sets timer to 100 msec.
ARM:TIMer DEFault Sets timer to 0.1 sec.
ARM:TIMer MINimum Sets timer to 1 msec.
ARM:TIMer MAXimum Sets timer to 99999.99 sec.
<clist>
Channel list — Specify one or more channels. Example:
ROUTe:SCAN (@101:110)
7702 scan list; slot 1, channels 1-10
<list>
List — Specify one or more numbers for a list. Example:
STATus:QUEue:ENABle (-110:-222)
Enable errors -110 through -222
Angle brackets < >
Angle brackets (< >) are used to denote a parameter type. Do
not include the brackets in the program message. For example:
RATio <b>
The <b> indicates a Boolean-type parameter is required.
Therefore, to enable channel ratio, you must send the com-
mand with the ON or 1 parameter as follows:
RATio ON
RATIO 1
Query commands
This type of command requests (queries) the presently programmed status. It is identified
by the question mark (?) at the end of the fundamental form of the command. Most com-
mands have a query form:
TRIGger:TIMer?
Queries the timer interval.
Most commands that require a numeric parameter (<n>) can also use the DEFault,
MINimum, and MAXimum parameters for the query form. These query forms are used to
determine the *RST default value and the upper and lower limits for the fundamental com-
mand. Examples are:
TRIGger:TIMer? DEFault
Queries the *RST default value.
TRIGger:TIMer? MINimum
Queries the lowest allowable value.
TRIGger:TIMer? MAXimum
Queries the largest allowable value.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...