Model 2790 SourceMeter
®
Switch System Reference Manual
17-13
AC current calibration
After the 700VAC at 1kHz point has been calibrated, the unit will prompt you for 100mA
at 1kHz. Follow these steps for AC current calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2790
as shown in
2.
Perform the calibration steps summarized in
. For each step:
•
Set the calibrator to the indicated current and frequency. Make sure the unit is
in OPERATE.
•
Press ENTER to complete each calibration step.
•
Allow the unit to complete each step before continuing.
Setting calibration dates and saving calibration
At the end of the calibration procedure, the instrument will display the CALIBRATION
COMPLETE message. Press ENTER to continue. The Model 2790 will prompt you to
enter the calibration date and the calibration due date.
Set these dates as follows:
1.
At the CAL DATE: prompt, use the left and right arrow keys and the range keys to
set the calibration date, then press ENTER.
2.
The unit will then prompt you to enter the next calibration due date with this
prompt: CAL NDUE:. Use the left and right arrow keys and the range keys to set
the calibration due date, then press ENTER.
3.
The unit will prompt you to save new calibration constants with this message:
SAVE CAL? YES. To save the new constants, press ENTER. If you do not want to
save the new constants, press the down range key to toggle to NO, then press
ENTER.
klqb
Calibration constants calculated during the present calibration procedure will
not be saved unless you choose the YES option. Previous calibration constants
will be retained if you select NO.
Table 17-7
AC current calibration summary
Calibration step
Calibrator current,
frequency
100mA at 1kHz
1A at 1kHz
2A at 1kHz
100.0000mA, 1kHz
1.000000A, 1kHz
2.000000A, 1kHz
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...