Model 2790 SourceMeter
®
Switch System Reference Manual
5-17
Source tracking
The Model 2790 provides source tracking to ensure that the most recently set source
amplitude is used in the selected ohms calculation. If the source output level changes, the
“m” constant for the calculation will change accordingly.
Source tracking is enabled automatically when the desired ohms function is selected. No
other actions are required by the user.
However, if desired, source tracking can be disabled or the source that is being tracked can
be changed. The :TRACk and :SOURCe commands are listed in
.
Math programming examples
Example #1
— The following command sequence performs the mX+b calculation for
channels 101 and 102 of the Model 7702. Keep in mind that after CALC:DATA? is sent,
the Model 2790 has to be addressed to talk to send the math result to the computer.
CALC:FORM MXB
' Select mX+b calculation.
CALC:KMAT:MMF 2
' Set 'm' factor to 2.
CALC:KMAT:MBF 100
' Set 'b' factor to 100.
CALC:STAT ON
' Enable math calculation.
ROUT:CLOS (@101)
' Close channel 101.
CALC:DATA?
' Read mX+b result for channel 101.
ROUT:CLOS (@102)
' Close channel 102.
CALC:DATA?
' Read mX+b result for channel 102.
Example #2
— The following command sequence configures channels 101 through 110
of the Model 7702 to perform the percent calculation when they are scanned.
CALC:FORM PERC,(@101:110)
' Select percent calculation.
CALC:KMAT:PERC 100,(@101:110)
' Set reference to 100.
CALC:STAT ON,(@101:*110)
' Enable math calculation.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...