5-10
Model 2790 SourceMeter
®
Switch System Reference Manual
For remote programming, the value +9.9e+37 is returned for an “OVRFLW” reading. The
value -9.9e+37 is returned for an “UDRFLW” reading.
m/X + b configuration
1.
Press SHIFT and then MATH to display the math menu.
2.
Press the RANGE
or
key to display “mX + B” and press ENTER to display
the present scale factor:
M: +1.000000
^
(factory default)
3.
Key in the scale factor value. The
and
keys control cursor position and
the
and
keys increment/decrement the digit value. To change range, place the
cursor on the multiplier and use the
and
keys (m =
×
0.001,
^
=
×
1,
K =
×
1000, and M =
×
1,000,000). With the cursor on the polarity sign,
the
and
keys toggle polarity.
4.
Press ENTER to enter the m value and display the offset (b) value:
b: +00.00000 m
(factory default)
5.
Key in the offset value.
6.
Press ENTER to enter the b value and display the one-character units designator:
UNITS: X
(factory default)
7.
Use the cursor keys and the
or
key if you wish to change the units designator.
The character can be any letter in the alphabet (A through Z), the degree symbol (°),
or the ohms symbol (
Ω
).
8.
Press ENTER. The MATH annunciator will turn on and the result of the calculation
will be displayed. Note that the calculation will be applied to all measurement
functions.
9.
To disable m/X + b, again press SHIFT and then MATH. The MATH annunciator
will turn off.
klqb
The result of the m/X+b calculation may be displayed in exponential notation.
For example, a displayed reading of +2.500E+03 R is equivalent to 2500 R
(2.5K R).
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...