Model 2790 SourceMeter® Switch System Reference Manual
G-17
:CODE
(:CALibration:PROTected:CODE)
Purpose
To program the calibration code or password so that you can perform the
Model 7751/7752/7753 calibration procedures.
Format
:cal:prot:code '<char_string>'
Parameters
Up to a 8-character string including letters and numbers.
Description
The
:CODE
command enables the Model 2790 calibration procedures
when performing these procedures over the bus. This command must be
sent to the unit before sending any other Model 7751/7752/7753 calibra-
tion command. The default calibration code is KI002790.
klqbp
The
:CODE
command should be sent only once before performing calibration.
Do not send
:CODE
before each calibration step.
The code parameter must be enclosed in single quotes.
Example
:CAL:PROT:CODE 'KI002790'
Send default code of KI002790.
:COUNt?
(:CALibration:PROTected:CARD1:COUNt?
(:CALibration:PROTected:CARD2:COUNt?
Purpose
To determine how many times a Model 7751/7752/7753 has been calibrated.
Format
:cal:prot:card1:coun?
:cal:prot:card2:coun?
Response
<n>
Calibration count.
Description
The
:CARD1:COUNt?
and :CARD2:COUNt?
commands allow you
to determine how many times a Model 7751/7752/7753 in Slot 1 or Slot 2
has been calibrated.
klqb
Use the
:COUNt?
command to help you monitor for unauthorized calibration
procedures.
Example
:CAL:PROT:CARD1:COUN?
Request card 1 calibration count.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...