17-24
Model 2790 SourceMeter
®
Switch System Reference Manual
5.
Calibrate step 0 with the following command:
:CAL:PROT:DC:STEP0
6.
Perform the entire remote comprehensive calibration procedure. See “
” earlier in this section.
7.
Connect the calibrator to the Model 2790 INPUT jacks as shown in
.
Select the front input jacks with the INPUTS switch.
8.
Set the calibrator to output a 1V RMS, 10Hz sine wave, then send the following
command:
:CAL:PROT:AC:STEP14 <Cal_voltage>
Here <Cal_voltage> is the actual 10Hz calibrator signal amplitude. If an error is
reported after performing this step, refer to
for troubleshooting.
9.
Connect the function generator to the Model 2790 INPUT jacks as shown in
10.
Set the generator to output a 1V RMS, 1kHz sine wave, then send the following
command:
:CAL:PROT:AC:STEP15 1E3
11.
Send the following commands to set calibration dates, save calibration, and lock
out calibration:
:CAL:PROT:DATE <yyyy>, <mm>, <dd>
:CAL:PROT:NDUE <yyyy>, <mm>, <dd>
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Calibration for firmware revision A03 and lower
Measuring function generator signal amplitude
The 3Hz function generator signal amplitude must be accurately measured using the digi-
tal multimeter listed in
. Proceed as follows:
1.
Connect the function generator output to the digital multimeter INPUT jacks. (See
for typical connections.)
2.
Turn on the function generator and multimeter. Allow a two-hour warm-up period
before measuring.
3.
Set the function generator to output a 1V RMS sine wave at 3Hz; measure and
record the signal amplitude.
Front panel manufacturing calibration (firmware revision A03 and lower)
1.
Install the shorted calibration extender board (see “
earlier in this section) in scanner card Slot 1 and select the rear inputs with the
INPUTS switch. Allow three minutes for thermal equilibrium.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...