Model 2790 SourceMeter
®
Switch System Reference Manual
5-15
Remote programming — math
Math commands
klqb
When measurements are performed, the readings are fed to other enabled
processing operations, including Math.
explains data flow and the
commands used to return Math results.
The commands to perform math calculations are listed in
. Details on these com-
mands follow the table.
klqb
Queries are not included in
. All the math commands are provided in
.
Table 5-2
Math commands
Commands
1
Description
Def
CALCulate[1]:FORMat <name> [, <clist>]
Select calculation; <name> = NONE, MXB
(mX+b), PERCent, RECiprocal (m/X+b),
S1Iohms, S2Iohms, S1Vohms, or S2Vohms.
PERC
CALCulate[1]:KMATh:MMFactor <NRf> [, <clist>]
Set “m” factor for mX+b and m/X+b;
<NRf> = -4294967295 to +4294967295.
1
CALCulate[1]:KMATh:MBFactor <NRf> [, <clist>]
Set “b” factor for mX+b and m/X+b;
<NRf> = -4294967295 to +4294967295.
0
CALCulate[1]:KMATh:MBFactor:TRACk <b>
Enable or disable source tracking
3
.
OFF
CALCulate[1]:KMATh:MBFactor:SOURce <name>
Selects source to be tracked; <name> = NONE,
S1ISource, S2ISource, S1VSource, S2VSource.
NONE
CALCulate[1]:KMATh:MUNits <char> [, <clist>]
Set units for mX+b and m/X+b; see “Setting
mX+b and m/X+b units”.
‘X’
CALCulate[1]:KMATh:PERCent <NRf> [, <clist>]
Set reference value for percent; <NRf> =
-4294967295 to +4294967295.
1
CALCulate[1]:KMATh:PERCent:ACQuire
Use input signal as reference value.
CALCulate[1]:STATe <b> [, <clist>]
Enable or disable calculation; <b> = ON or OFF.
(Note 2)
CALCulate[1]:DATA[:LATest]?
Return last result of calculation.
CALCulate[1]:DATA:FRESh?
Return last “fresh” result of calculation.
Channel list parameter:
7702 <clist> = (@SCH)
where:
S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
7751/7752/7753 <clist> (DCV,
Ω
2,
Ω
4) = (@Mn) (see
).
1. The <clist> parameter is used to configure one or more 7702 channels for a scan.
2. *RST default is ON, SYSTem:PRESet is OFF.
3. Selecting an ohms calculation enables source tracking. Selecting any other math function disables source tracking.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...