Model 2790 SourceMeter
®
Switch System Reference Manual
4-5
Manual ranging
The range is selected by specifying the expected reading as an absolute value using the
<n> parameter for the appropriate :RANGe command. The Model 2790 will then go to the
most sensitive range for that expected reading. For example, if you expect a reading of
approximately 3V, let the parameter (<n>) equal 3 to select the 10V range.
Auto ranging
The :RANGe:AUTO command is coupled to the command to select range manually
(:RANGe <n>). When auto range is enabled, the parameter value for :RANGe <n>
changes to the automatically selected range value. When auto range is disabled, the instru-
ment remains at the selected range. When a valid :RANGe <n> command is sent, auto
ranging disables.
Range programming examples
Example #1
— The following commands select range for DCV,
Ω
2, and DCI:
VOLT:RANG 0.5
' Select 1V range for DCV.
RES:RANG 2e3
' Select 10k
Ω
range for
Ω
2.
CURR:RANG 0.1
' Select 100mA range for DCI.
Example #2
— The following command sequence configures channel 101 of the
Model 7702 to select the 10VDC range when it is scanned.
FUNC 'VOLT',(@101)
' Set 101 for DCV function.
VOLT:RANG 1.5,(@101)
' Set 101 for 10V range.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...