17-8
Model 2790 SourceMeter
®
Switch System Reference Manual
3.
When the unit is finished with short-circuit calibration, it will display the following
prompt:
OPEN CIRCUIT
4.
Remove the calibration short and press ENTER. During this phase, the
CALIBRATING message will be displayed.
klqb
Be sure to minimize movement near front input terminals. Excessive movements
can cause capacitive coupling errors, which could affect calibration accuracy.
DC volts calibration
After the front panel short and open procedure, the unit will prompt you for the first DC
voltage: +10V. Perform the following:
1.
Connect the calibrator to the Model 2790 as shown in
. Wait three min-
utes to allow for thermal equilibrium before proceeding.
Figure 17-2
Connections for DC volts and ohms calibration
klqb
Although 4-wire connections are shown, the sense leads are connected and dis-
connected at various points in this procedure by turning calibrator external
sense on or off as appropriate. If your calibrator does not have provisions for
turning external sense on and off, disconnect the sense leads when external sens-
ing is to be turned off and connect the sense leads when external sensing is to be
turned on.
F
!
F
Model 2790
SourceMeter Switch System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
RS-232
GPIB
NOTE
Use shielded low-thermal cables to
minimize noise. Enable and disable
calibrator external sense as needed.
Model 2790
DC Voltage and Resistance Calibrator
SENSE HI INPUT HI
SENSE LO
INPUT LO
SENSE HI
SENSE LO
OUTPUT
HI
OUTPUT
LO
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...