D-2
Model 2790 SourceMeter
®
Switch System Reference Manual
Measurement considerations
Low-level voltage measurements made using the Model 2790 can be adversely affected by
various types of noise or other unwanted signals that can make it very difficult to obtain
accurate voltage readings. Some of the phenomena that can cause unwanted noise include
thermoelectric effects (thermocouple action), source resistance noise, magnetic fields, and
radio frequency interference. The following paragraphs discuss the most important of
these effects and ways to minimize them.
klqb
For comprehensive information on low-level measurements, see the “Low Level
Measurements” handbook, which is available from Keithley.
Thermoelectric potentials
Thermoelectric potentials (thermal EMFs) are small electric potentials generated by
differences in temperature at the junction of dissimilar metals. The following paragraphs
discuss how such thermals are generated and ways to minimize their effects.
Thermoelectric coefficients
, the magnitude of thermal EMFs generated depends on the particu-
lar materials involved. Best results are obtained with clean copper-to-copper connections
as indicated in the table.
Table D-1
Material thermoelectric coefficients
Material
Thermoelectric potential
Copper-Copper
Copper-Silver
Copper-Gold
Copper-Cadmium/Tin
Copper-Lead/Tin
Copper-Kovar
Copper-Silicon
Copper-Copper Oxide
0.2µV/°C
0.3µV/°C
0.3µV/°C
0.3µV/°C
1–3µV/°C
40µV/°C
400µV/°C
1000µV/°C
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...